VD Series AC/DC High Voltage Dividers Udey Test Kits
VD Series AC/DC High Voltage Dividers Udey Test Kits
UDEY series ‘VD’ consists of two parts: High voltage divider and digital meter. The top terminal of the voltage divider is connected to high voltage output under test / calibration. The voltage divider is connected to digital meter by measuring cable provided.
Both, the voltage divider and digital meter are stored in transportable light weight single aluminum cabinet with handle.
Machine Feature
- Digital display
- One tester for AC and DC
- Saves set up time
- Machine Technical Specification
Measuring range | AC [50HZ] DC | 50 kv 50 kv | 100 kv 100kv | 200 kV 200 kV |
Accuracy | AC [50HZ] DC | 1.50% 1.00% | 1.5% 1.0% | 1.5% 1.0% |
Resistance | R [ M ] C [pF] | 640 300 | 1280 150 | 2560 75 |
Divide ratio | K | 1000 | 1000 | 1000 |
Disply | LCD | LCD | LCD | |
Maximum error | AC DC | 750V 500V | 1500V 1000V | 3000V 2000V |
Meter input resistance | 10 MΩ | 10 MMΩ | 10 MΩ | |
Dimensions [mm] | 200x200x520 | 200x200x850 | 200x200x1160 | |
Weight [kg] | 9.5 | 13.5 | 18.5 | |
Ambient | Temperature Relative humidity | 20 ± 15º C To 80 % non condensing |
High Voltage Divider
The AC/DC High Voltage Dividers, VD series consists of two main parts which are the digital display meter and the High voltage divider. A high voltage output under calibration or test is connected from the top terminal of the voltage divider. The High Voltage Divider is connected to a digital meter by a provided measuring cable. The setup time is saved in the high voltage divider for accurate measurement and it has an LCD digital display to show the results. Both, the digital meter and the high voltage divider are stored in a transportable single lightweight cabinet made of aluminum with handle. The high voltage divider device has one tester for the alternating type current and another tester for the direct type current.
The measuring range of the device starts from a 50 kilo voltage and up to 100 kilo voltage for the alternating current measurement and the same range is there for the direct current measurement. The weight of this tester ranges from 9.5 Kilograms up to 18.5 Kilograms. The great advantage of this device is the high level of accuracy, so the results shown by this device are reliable for decision making because the inaccuracy error is as low as 1% up to 1.5%.
Explore the comprehensive range of Udey test kits click here.